Emergency Medicine Journal 2005;22:177-179; doi:10.1136/emj.2003.011619
© 2005 BMJ Publishing Group Ltd and the College of Emergency Medicine.

ORIGINAL ARTICLE

Capillary versus venous bedside blood glucose estimations

R Boyd, B Leigh, P Stuart

Emergency Department, Lyell McEwin Hospital, Haydown Road, Elizabeth Vale 5112, South Australia

Correspondence to:
Correspondence to:
Dr R Boyd
Emergency Department, Lyell McEwin Hospital, Haydown Road, Elizabethvale 5112, South Australia; russell.boyd{at}nwahs.sa.gov.au

Objectives: To determine the mean difference and correlation between capillary and venous bedside glucose estimation in comparison to laboratory blood glucose analysis in emergency department (ED) patients.

Methods: Blood glucose levels were synchronously analysed using a bedside blood glucometer on capillary and venous derived samples from consenting ED patients aged >12 years. The venous sample was sent for comparative testing using a laboratory based multichannel analyser. Mean difference and correlation coefficients were determined.

Results: A total of 20 subjects (aged 13–88 years) were enrolled, with 100% data capture. The mean laboratory glucose was 7.075 mmol/l. The mean capillary blood glucose was 7.66 mmol/l (mean difference compared with mean laboratory glucose 0.58 mmol/l; 95% confidence interval 0.3 to 0.9). The mean venous derived blood glucometer glucose was 7.99 mmol/l (mean difference compared with mean laboratory glucose 0.91 mmol/l; 95% CI 0.6 to 1.2). The correlation coefficient for the laboratory blood glucose versus the capillary blood glucometer glucose was 0.97 mmol/l (p<0.001). The correlation coefficient for the laboratory blood glucose and the venous blood glucometer glucose was 0.96 (p<0.001). Variation occurred between the glucometer and the laboratory blood glucose results.

Conclusions: There is a small but significant difference in the blood glucose results analysed on a bedside glucometer when the samples are taken from capillary or venous sources. Although good correlation is the norm between venous and capillary derived samples, caution must be exercised in accepting the results as equivalent or using either as substitutes for a laboratory blood glucose result.

Abbreviations: ED, emergency department; IV, intravenous

Keywords: blood glucose; venous; capillary; measurement; bedside


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

This article has been cited by other articles:

  • Shearer, A., Boehmer, M., Closs, M., Dela Rosa, R., Hamilton, J., Horton, K., McGrath, R., Schulman, C. (2009). Comparison of Glucose Point-of-Care Values With Laboratory Values in Critically Ill Patients. Am J Crit Care 18: 224-230 [Abstract] [Full Text]  
  • Mahoney, J., Ellison, J. (2007). Assessing the Quality of Glucose Monitor Studies: A Critical Evaluation of Published Reports. Clin. Chem. 53: 1122-1128 [Abstract] [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

 

The journal is co-owned by and the official journal of College of Emergency Medicine

Official journal of British Association for Immediate Care: BASICS, Faculty of Pre-Hospital Care, Irish Society for Immediate Care and Swedish Society for Emergency Medicine: SweSEM

Emergency Medicine Jobs

Emergency Medicine Jobs