ORIGINAL ARTICLES
Comparison of the Simplify D-dimer assay performed at the bedside with a laboratory-based quantitative D-dimer assay for the diagnosis of pulmonary embolism in a low prevalence emergency department population
1 Department of Emergency Medicine, Carolinas Medical Center, Charlotte, North Carolina, USA
2 Department of Pathology, Carolinas Medical Center, Charlotte, North Carolina, USA
Correspondence to:
Dr J A Kline, Emergency Medicine Research, Department of Emergency Medicine, Carolinas Medical Center, PO Box 32861, Charlotte, North Carolina 28323-2861, USA; jkline{at}carolinas.org
Background: The immunofiltration D-dimer assay could allow point-of-care testing for pulmonary embolism (PE). A study was undertaken to compare a clinician-performed qualitative D-dimer assay with the automated quantitative D-dimer test.
Methods: A prospective observational study was conducted from January to October 2005 at an urban academic emergency department (ED). 1193 patients of mean (SD) age 47 (16) years (66% female) were enrolled. The study protocol combined pretest probability estimation, D-dimer testing by both a qualitative immunochromatographic assay (Simplify) performed at the point of care by 192 different clinicians and a quantitative D-dimer test performed in a CLIA-certified laboratory. The criterion standard was image-proven PE or deep venous thrombosis within 45 days after enrolment. To test interobserver agreement for the qualitative assay, two blinded observers independently read 841 Simplify cartridges.
Results: Of 1193 patients enrolled, 45 were PE+ (3.8%, 95% CI 2.8% to 5.0%). Qualitative results were available for 1169 (98%) and quantitative results were available for 1136 (95%). Comparison of the qualitative and quantitative D-dimer tests gave the following results: sensitivity 91% (95% CI 78% to 98%) vs 93% (95% CI 80% to 98%); specificity 57% (95% CI 54% to 60%) vs 57% (95% CI 54% to 60%); likelihood ratio negative 0.16 (95% CI 0.06 to 0.37) vs 0.13 (95% CI 0.05 to 0.35). The weighted Cohens
for interpretation of the qualitative assay was 0.69 (95% CI 0.63 to 0.76).
Conclusions: In this very low-risk ED population, a qualitative D-dimer assay performed at the point of care had similar diagnostic accuracy to the quantitative D-dimer test. Interobserver agreement for the qualitative test was good.
This article has been cited by other articles:
-
Gupta, R. T., Kakarla, R. K., Kirshenbaum, K. J., Tapson, V. F.
(2009). D-Dimers and Efficacy of Clinical Risk Estimation Algorithms: Sensitivity in Evaluation of Acute Pulmonary Embolism. Am. J. Roentgenol.
193: 425-430
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.