Near-patient testing of potassium levels using arterial blood gas analysers: can we trust these results?

R J P José, J Preller

Correspondence to:
Dr R J P José, Department of Intensive Care Medicine, Broomfield Hospital, Court Road, Broomfield CM1 7ET, UK; rjpj{at}ananzi.co.za

Background: Near-patient testing allows rapid availability of results to enable prompt decision-making. Potassium abnormalities are common in acutely ill patients and can be associated with life-threatening complications. At times there is uncertainty whether clinical decisions can be based on the potassium result obtained from arterial blood gas (ABG) analysers or if laboratory values should be awaited.

Objectives: To determine the opinion of clinicians regarding the use of blood gas analysers to measure potassium and to determine the level of agreement between blood gas analyser and laboratory measurements of potassium in arterial blood samples.

Method: Survey of 64 doctors using a questionnaire and a retrospective comparative study of 529 paired results of ABG and arterial laboratory measurements of potassium in 121 critically ill patients.

Results: 51.6% of the doctors would wait for laboratory confirmation and 48.4% would base clinical decisions on results obtained from the blood gas analyser. The difference between the means of potassium values from the two methods is 0.03 mmol/l (95% CI 0.011 to 0.056; p = 0.0041). The 95% limits of agreement were from –0.485 mmol/l (95% CI –0.524 to –0.447) to 0.551 mmol/l (95% CI 0.513 to 0.590). 95% of the results fell within the difference limits of 0.5 mmol/l.

Conclusions: Most clinicians still await laboratory confirmation of results obtained from blood gas analysers but in this setting there is sufficient agreement between the results obtained from the authors' blood gas analyser and a laboratory analyser to enable effective clinical desisions to be made.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Primary survey
Steve Goodacre
Emerg. Med. J. 2008 25: 467. [Extract] [Full Text] [PDF]

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

 

The journal is co-owned by and the official journal of College of Emergency Medicine

Official journal of British Association for Immediate Care: BASICS, Faculty of Pre-Hospital Care, Irish Society for Immediate Care and Swedish Society for Emergency Medicine: SweSEM

Emergency Medicine Jobs

Emergency Medicine Jobs