ACCURACY OF PULSE OXIMETRY

Sophia is aware that nail varnish can spell trouble with pulse oximetry, but what of artificial acrylic nails? A trial reported in Resuscitation (2007;74:75–82) compared two different pulse oximeters on intensive care patients to determine whether false nails affected the readings. Each patient had a pulse oximeter applied to a natural nail and to an acrylic nail, and the results were compared with an arterial blood gas measurement of oxygen saturations. The two oximeters tested emitted slightly different wavelengths of light and this may explain why one of them gave inaccurate readings when applied to false nails. The authors recommend that if false nails are in place, the oximeter probe be placed at 90° to the usual position so that the light does not pass through the nail (the alternative is spending 30 minutes soaking the nail off in acetone).

COMORBIDITY AND CARDIAC ARREST

A US study of more than 1000 patients reports the influence of chronic comorbidity on survival after out-of-hospital VF arrest due to heart disease. The odds of survival to …

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.