Table 2

Sensitivity and specificity of BIS monitoring for predicting poor neurological outcome after cardiac arrest

StudyTPTNFPFNSensitivity (95% CI)Specificity (95% CI)Cut-off BIS value used in study
Leary 2010417320.67 (0.22–0.96)0.85 (0.62–0.97)23
Seder 20094331270.86 (0.73–0.94)0.94 (0.80–0.99)22
Selig 2014617210.86 (0.42–1.00)0.89 (0.67–0.99)40
Shibata 200514140170.45 (0.27–0.64)1.00 (0.77–1.00)40
Stammet 200933030.50 (0.12–0.88)1.00 (0.29–1.00)0
POOLED RESULTS0.70 (0.49–0.85)0.93 (0.83–0.97)
  • TP, true positives; TN, true negatives; FP, false positives; FN, false negatives.