Comparison of a novel rectilinear biphasic waveform with a damped sine wave monophasic waveform for transthoracic ventricular defibrillation. ZOLL Investigators

J Am Coll Cardiol. 1999 Nov 1;34(5):1595-601. doi: 10.1016/s0735-1097(99)00363-0.

Abstract

Objectives: We compared the efficacy of a novel rectilinear biphasic waveform, consisting of a constant current first phase, with a damped sine wave monophasic waveform during transthoracic defibrillation.

Background: Multiple studies have shown that for endocardial defibrillation, biphasic waveforms have a greater efficacy than monophasic waveforms. More recently, a 130-J truncated exponential biphasic waveform was shown to have equivalent efficacy to a 200-J damped sine wave monophasic waveform for transthoracic ventricular defibrillation. However, the optimal type of biphasic waveform is unknown.

Methods: In this prospective, randomized, multicenter trial, 184 patients who underwent ventricular defibrillation were randomized to receive a 200-J damped sine wave monophasic or 120-J rectilinear biphasic shock.

Results: First-shock efficacy of the biphasic waveform was significantly greater than that of the monophasic waveform (99% vs. 93%, p = 0.05) and was achieved with nearly 60% less delivered current (14 +/- 1 vs. 33 +/- 7 A, p < 0.0001). Although the efficacy of the biphasic and monophasic waveforms was comparable in patients with an impedance < 70 ohms (100% [biphasic] vs. 95% [monophasic], p = NS), the biphasic waveform was significantly more effective in patients with an impedance > or = 70 ohms (99% [biphasic] vs. 86% [monophasic], p = 0.02).

Conclusions: This study demonstrates a superior efficacy of rectilinear biphasic shocks as compared with monophasic shocks for transthoracic ventricular defibrillation, particularly in patients with a high transthoracic impedance. More important, biphasic shocks defibrillated with nearly 60% less current. The combination of increased efficacy and decreased current requirements suggests that biphasic shocks as compared with monophasic shocks are advantageous for transthoracic ventricular defibrillation.

Publication types

  • Clinical Trial
  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Aged
  • Arrhythmias, Cardiac / therapy*
  • Electric Countershock / methods*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Treatment Outcome