Article Text

Download PDFPDF
Agreement between mathematically arterialised venous versus arterial blood gas values in patients undergoing non-invasive ventilation: a cohort study
  1. Anne-Maree Kelly1,2,
  2. Sharon Klim1,
  3. Stephen E Rees3
  1. 1Joseph Epstein Centre for Emergency Medicine Research at Western Health, Melbourne, Australia
  2. 2The University of Melbourne, Melbourne, Australia
  3. 3Respiratory and Critical Care Group (RCARE), Center for Model-Based Medical Decision Support, Institute for Health Science and Technology, Aalborg University, Aalborg, Denmark
  1. Correspondence to Professor Anne-Maree Kelly, JECEMR, Sunshine Hospital, Furlong Road, St Albans 3021, Australia; anne-maree.kelly{at}wh.org.au

Abstract

Background Blood gas analysis is important for assessment of ventilatory function. Traditionally, arterial analysis has been used. A method for mathematically arterialising venous blood gas values has been developed. Our aim was to validate this method in patients undergoing non-invasive ventilation (NIV) in an emergency department (ED).

Materials and methods This post hoc substudy of a prospective cohort study included adult patients undergoing NIV for acute respiratory compromise. When arterial blood gas analysis was required for clinical purposes, a venous sample was also drawn. Mathematically arterialised values were calculated independent of arterial values. Primary outcome of interest was agreement between mathematically arterialised venous and arterial values for pH and pCO2. Bland-Altman agreement plot analysis was used.

Results Eighty sample-pairs (58 patients) were studied. Mean difference for arterial pH (actual-calculated) was 0.01 pH units (95% limits of agreement: −0.04, 0.06). Mean difference for pCO2 (actual-calculated) was −0.06 kPa (95% limits of agreement: −1.34, 1.22).

Conclusions For patients undergoing NIV in an ED, agreement between mathematically arterialised venous values and arterial values was close for pH but only moderate for pCO2. Depending on clinician tolerance for agreement, this method may be a clinically useful alternative to arterial blood gas analysis in the ED.

  • Emergency Department
  • Respiratory
  • Ventilation, Non invasive

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.