The incidence of airway haemorrhage in manual versus mechanical cardiopulmonary resuscitation
  1. Stephen Edward Asha1,2,
  2. Sarah Doyle2,
  3. Glenn Paull3,
  4. Victar Hsieh2,3
  1. 1 Emergency Department, St George Hospital, Kogarah, New South Wales, Australia
  2. 2 St George and Sutherland Clinical School, University of New South Wales, Sydney, New South Wales, Australia
  3. 3 Cardiology Department, St George Hospital, Kogarah, New South Wales, Australia
  1. Correspondence to Associate Professor (Conjoint) Stephen Edward Asha, Emergency Department, St George Hospital, Kogarah, NSW 2217, Australia; stephen.asha{at}health.nsw.gov.au

Abstract

Objective The aim of this study was to compare the incidence of airway haemorrhage between participants who received manual cardiopulmonary resuscitation (CPR) and those who had received mechanical CPR using the LUCAS device.

Methods A retrospective cohort study was conducted by means of a medical chart review. All non-traumatic cardiac arrest patients that presented to the ED, from May 2014 to February 2018, were recruited. The groups were stratified according to those who had the majority of CPR performed using the LUCAS and those who had the majority of CPR performed manually. The primary outcome was the proportion of participants with airway haemorrhage, defined as blood observed in the endotracheal tube, pharynx, trachea or mouth, and documented in the doctor or nursing notes. Logistic regression analysis was performed to adjust for confounders.

Results 12 of 54 (22%) participants in the majority LUCAS CPR group had airway haemorrhage, compared with 20 of 215 (9%) participants in the majority manual CPR group, a difference of 13% (95% CI 3% to 26%, p=0.02). The unadjusted odds for developing airway haemorrhage in the majority LUCAS CPR group was 2.8 (95% CI 1.3 to 6.1). After adjusting for confounders, the odds for developing airway haemorrhage in the majority LUCAS CPR group was 2.5 (95% CI 1.1 to 5.7).

Conclusions The LUCAS mechanical CPR device is associated with a higher incidence of airway haemorrhage compared with manual CPR. Limitations in the study design mean this conclusion is not robust.

  • airway
  • cardiac arrest
  • chest
  • resuscitation
  • Trauma

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Contributors SEA designed the study, supervised the conduct of the study, performed chart review and statistical analysis, revised the manuscript. SD performed the literature search, chart review, collated all data, and drafted and revised the manuscript. VH and GP reviewed the study design, identified participants and revised the manuscript. SEA takes responsibility for the manuscript.

  • Funding The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

  • Competing interests None declared.

  • Patient consent for publication Not required.

  • Ethics approval Permission for this study was granted by the South Eastern Sydney Local Health District Human Research Ethics Committee, Sydney, Australia; reference number 16/382(LNR/16/POWH/717).

  • Provenance and peer review Not commissioned; externally peer reviewed.

  • Data availability statement All data relevant to the study are included in the article or uploaded as supplementary information.