PT - JOURNAL ARTICLE AU - P Leonard AU - T F Beattie AU - P S Addison AU - J N Watson TI - Standard pulse oximeters can be used to monitor respiratory rate AID - 10.1136/emj.20.6.524 DP - 2003 Nov 01 TA - Emergency Medicine Journal PG - 524--525 VI - 20 IP - 6 4099 - http://emj.bmj.com/content/20/6/524.short 4100 - http://emj.bmj.com/content/20/6/524.full SO - Emerg Med J2003 Nov 01; 20 AB - Background: One of the most important limitations of standard pulse oximeters is the inability to detect changes in respiratory rate until oxygenation is affected. This study sought to determine if analysis of the plethysmogram by wavelet transforms would enable the determination of changes in respiratory rate at an earlier stage. Methods: Ten healthy adult volunteers were monitored, breathing at baseline and predetermined respiratory rates, using a standard pulse oximeter. Photo-plethysmograms captured in an attached lap top computer were then analysed using wavelet transforms. Results: Determination of baseline respiratory rate and subsequent changes including apnoea were easily identified. Comment: Wavelet transforms permit the accurate determination of respiratory rate by a standard pulse oximeter.