PT - JOURNAL ARTICLE AU - Erez Dvir AU - Danny Epstein AU - Baruch Berzon AU - Aeyal Raz AU - Amit Lehavi TI - Differences in the thermal properties and surface temperature of prehospital antihypothermia devices: an in vitro study AID - 10.1136/emermed-2020-211057 DP - 2021 Nov 03 TA - Emergency Medicine Journal PG - emermed-2020-211057 4099 - http://emj.bmj.com/content/early/2021/11/05/emermed-2020-211057.short 4100 - http://emj.bmj.com/content/early/2021/11/05/emermed-2020-211057.full AB - Background Preventing and treating hypothermia in prehospital settings is crucial. Several products have been developed to prevent heat loss and actively warm patients in prehospital settings. We compared the efficacy and the surface temperature of different antihypothermia products, using a fluid-based model at two ambient temperatures.Methods We tested five active (Blizzard Heat with active pads, Ready-Heat, Ready-Heat-II, Hypothermia Prevention and Management Kit (HPMK), Bair Hugger) and five passive (Blizzard Heat, Heat Reflective Shell, sleeping bag, ‘space blanket’, wool blanket) antihypothermia products. A torso model consisting of four 8 L bags of fluid preheated to 36°C±0.5°C (97±0.5°F) was used to compare the devices’ performances at 20°C (68°F) and 8°C (46°F). Inner and surface temperatures were recorded for up to 480 min.Results We found significant differences in heat loss in fluid bags among the tested devices at both temperatures (p<0.001). At 20°C, only HPMK and Ready-Heat-II increased the inner temperature for 480 min while Blizzard Heat with active pads prevented heat loss. Ready-Heat prevented heat loss for 90 min. All the other devices did not prevent heat loss beyond 30 min. At 8°C, none of the products heated the model. Bair Hugger, HPMK, Ready-Heat II and sleeping bag prevented heat loss for 30 min. At 60, 90 and 120 min HPMK, Ready-Heat II and Bair Hugger were the most effective. Over 480 min, Bair Hugger was most effective, with a heat loss of 2.3°C±0.4°C. The surface temperature exceeded 44°C (111°F) for all the exothermic warming devices when used for a prolonged period of time.Conclusion At 20°C, HPMK and Ready-Heat-II increased fluid temperature in the model, while the other devices decreased heat loss. At 8°C, none of the tested devices increased the temperature. However, active heating devices prevented heat loss slightly better than passive methods. A protective insulation layer should be used with all active heating blankets.Data are available on reasonable request.