Table 3

Results from diagnostic accuracy studies of POCT hs-cTn assays at presentation for the diagnosis of AMI

POC assayAUC
(95% CI)
Comparator assayAUC
(95% CI)
PatientsAMI rate
PATHFAST POC hs-cTnI7
(plasma)
0.91
(0.89 to 0.93)
cTnI-Architect
(fresh serum or plasma)
0.90
(0.87 to 0.92)
1279134 (20%)
i-STAT TnI-Nx20 *
(plasma)
0.97
(0.96 to 0.99)
cTnI-Architect
(plasma)
0.97
(0.95 to 0.99)
35457 (16%)
Minicare POC hs-cTnI9
(whole blood)
0.88
(0.83 to 0.94)
cTnI-Architect
(serum or plasma)
0.91
(0.87 to 0.95)
45072 (16%)
I-Stat POC cTnI0.88
(0.82 to 0.94)
Triage True POC hs-cTnI6
(plasma)
0.95
(0.93 to 0.96)
cTnT Elecsys
(serum or plasma)
0.94
(0.93 to 0.96)
1261178 (14%)
cTnI-Architect
(serum or plasma)
0.92
(0.90 to 0.93)
  • *Analytical studies of this assay are pending.

  • AMI, acute myocardial infarction; AUC, area under the curve; hs-cTnI, high-sensitivity cardiac troponin I; POCT, point-of-care test.