Handheld metal detector localization of ingested metallic foreign bodies: accurate in any hands?

Arch Pediatr Adolesc Med. 1999 Aug;153(8):853-7. doi: 10.1001/archpedi.153.8.853.

Abstract

Objective: To demonstrate the accuracy of handheld metal detectors (HHMDs) for identification and localization of ingested metallic foreign bodies when used by experienced and inexperienced investigators.

Design: Prospective study comparing HHMD scanning with radiography.

Setting and patients: A consecutive sample of all eligible patients (N = 176) presenting to the emergency departments of Children's Medical Center of Dallas, Dallas, Tex, and Children's Hospital of the King's Daughters, Norfolk, Va, who were known or suspected to have ingested a metallic foreign body.

Intervention: Each patient underwent HHMD scanning and radiographic evaluation.

Main outcome measures: Statistical evaluation compared HHMD scanning with radiography and experienced vs inexperienced investigator HHMD scanning to determine the accuracy of the screening tool and investigators.

Results: Experienced investigators performed HHMD scans on 140 subjects; inexperienced investigators scanned all subjects. Disease was defined as a foreign body in the esophagus on radiograph. The 3 experienced investigators demonstrated sensitivity of 100% (95% confidence interval [CI], 94%-100%), specificity of 92.4% (95% CI, 84.2%-97.2%), positive predictive value (PPV) of 90.9% (95% CI, 81.3%-96.6%), and negative predictive value (NPV) of 100% (95% CI, 95.1%-100%); the proportion correct was 95.7% (95% CI, 90.8%-98.4%). The inexperienced investigators demonstrated sensitivity of 95.7% (95% CI, 88.0%-99.1%), specificity of 81.0% (95% CI, 72.1%-88.0%), PPV of 77.0% (95% CI, 66.8%-85.4%), and NPV of 96.6% (95% CI, 90.4%-99.3%); the proportion correct was 86.9% (95% CI, 80.9%-91.5%). The McNemar test demonstrated no statistically significant difference between HHMD scanning by experienced vs inexperienced investigators.

Conclusions: Handheld metal detector scanning is an accurate, inexpensive, radiation-free screening tool and should be used for evaluation of patients suspected of ingesting coins and coinlike foreign bodies.